Knee Surg Relat Res 2016 Mar; 28(1): 39-45  https://doi.org/10.5792/ksrr.2016.28.1.39
Two- to Four-Year Follow-up Results of Total Knee Arthroplasty Using a New High-Flexion Prosthesis
Man Soo Kim, MD, In Jun Koh, MD, Sung Won Jang, MD, Neung Han Jeon, MD, and Yong In, MD*

Department of Orthopaedic Surgery, Seoul St. Mary's Hospital, The Catholic University of Korea College of Medicine, Seoul, Korea.

Correspondence to: Correspondence to: Yong In, MD. Department of Orthopaedic Surgery, Seoul St. Mary's Hospital, The Catholic University of Korea College of Medicine, 222 Banpo-daero, Seocho-gu, Seoul 06591, Korea. Tel: +82-2-2258-2838, Fax: +82-2-535-9834, iy1000@catholic.ac.kr
Received: September 13, 2015; Revised: December 2, 2015; Accepted: December 15, 2015; Published online: February 29, 2016.
© Korean Knee Society. All rights reserved.

This is an Open Access article distributed under the terms of the Creative Commons Attribution Non-Commercial License (http://creativecommons.org/licenses/by-nc/4.0) which permits unrestricted non-commercial use, distribution, and reproduction in any medium, provided the original work is properly cited.
Abstract

Purpose

The purpose of this study was to evaluate minimum 2-year follow-up results of total knee arthroplasty (TKA) performed using a new high-flexion prosthesis design (LOSPA).

Materials and Methods

The 2- to 4-year results of 191 consecutive TKAs (177 patients) with the LOSPA posterior-stabilized prosthesis were evaluated. The patients were assessed clinically and radiographically using the Knee Society scoring system (KSS) and the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC).

Results

The mean range of motion (ROM) increased significantly from 117.4° (range, 75° to 140°) preoperatively to 126.7° (range, 80° to 144°) postoperatively (p<0.001). The mean KSS and WOMAC scores improved significantly from 121.4 (range, 42 to 185) and 56.1 (range, 23 to 88) preoperatively to 174.0 (range, 130 to 200) and 16.4 (range, 0 to 85) postoperatively, respectively (both, p<0.001). One knee required revision for deep infection. No knee had aseptic loosening or osteolysis. Radiolucent lines were noted in 15 knees (7.9%).

Conclusions

The new high-flexion total knee prosthesis resulted in no early aseptic loosening of the component and improved postoperative ROM comparable to other high-flexion TKA prostheses at 2- to 4-year follow-ups.

Keywords: Knee, Arthroplasty, Range of motion, Prosthesis design, Outcome assessment
Introduction

Total knee arthroplasty (TKA) has been widely accepted as an effective treatment for end stage osteoarthritis demonstrating satisfactory outcomes and survivorship. TKA is advantageous in relieving pain and improving knee function1). Still, restoration of normal range of motion (ROM) after TKA remains as a challenge; in general, the ROM of the knee after conventional TKA (range, 110° to 120°) is less than that of a normal knee2). In particular, squatting, sitting cross-legged, and kneeling with deep flexion of the knee joint that involve >125° of knee flexion are often required during daily living in Asian culture3). Thus, high-flexion prostheses were developed to meet such demands4).

High-flexion TKA is designed to achieve >125° of knee flexion5). The femoral component of the high-flexion design has a 2?4 mm extended and thicker posterior condyle that does not necessitate additional bone cut to maintain the contact area during deep knee flexion67). The increased contact area helps posterior femoral translation. In addition, the larger posterior condyle of the femoral component decreases the risk of impingement between the posterior cortex of the femur and the polyethylene insert8). The new prosthesis has a modified cam and post mechanism with increased jump distance to avoid dislocation under deep flexion angles9). It also has a longer trochlear groove for improved patellar tracking. The anterior recess in the tibial polyethylene insert reduces impingement on the patella and the patellar tendon during deep knee flexion.

The new high-flexion concept LOSPA knee system (Corentec, Seoul, Korea) has recently become available. It is designed to achieve ligament isometry from 0° to 90° using a single radial axis of the femoral component. The single radius of the femoral component contributes to stability in the mid-flexed state10). Cam-post engagement occurs at 75°, which prevents unnecessary cam-post contact during level walking and has an advantage of posterior translation of the femoral condyle. The high-flexion design with more rounded femoral contour and deepened patellar groove facilitates deep flexion by reducing joint capsule overstuffing. The elongated and laterally angled patellar groove is constructed to produce good patellar tracking. This prosthesis has a 10-mm thick posterior condyle of the femoral component that replaces the 10-mm posterior femoral bone cut, which contributes to increased surface contact and bone support through a relatively larger posterior radius than the conventional design (Fig. 1). Compared to Nexgen LPS-flex (Zimmer, Warsaw, IN, USA), another high-flexion design implant, that requires 12-mm posterior femoral bone cutting, LOSPA prosthesis is more advantageous for bone preservation.

Numerous studies have compared the high-flexion and conventional designs. Some studies demonstrated the superiority of the high-flexion prosthesis with regard to the postoperative ROM and function411) while others showed the opposite712). A concern has been raised that high-flexion TKA accelerates early aseptic loosening compared to conventional TKA due to the increased shear force between the femoral component and the femur during high flexion of the knee1314). Cho et al.13) and Han et al.14) noted a high rate of aseptic loosening of the femoral component in their patients after high-flexion TKA. In this study, we retrospectively evaluated the clinical and radiographic results of TKA performed using the newly introduced LOSPA knee system for a minimum 2-year follow-up. We examined the following two hypotheses: 1) the ROM of the knee with the LOSPA prosthesis would be comparable to that with other high-flexion concept prostheses of the recent literatures in Korean patients and 2) the implantation of the LOSPA prosthesis would be stable without aseptic loosening at minimum 2- year follow-up.

Materials and Methods

A total of 198 patients (215 knees) underwent TKA with the posterior-stabilized (PS) LOSPA total knee prosthesis from May 2011 to August 2013 at our institution. Of these, we retrospectively reviewed 191 knees of 177 patients (88%) who were available for a minimum 2-year follow-up. Inclusion criteria were patients with indications for primary TKA: Kellgren-Lawrence classification grade 3?4 severe osteoarthritis, osteonecrosis15), and rheumatoid arthritis. Exclusion criteria were contraindications of TKA such as knee joint infection. Patient demographics are summarized in Table 1. The mean follow-up period was 2.6 years (range, 2 to 4 years).

All surgeries were performed by a senior author (In) using the PS LOSPA knee system. All procedures were performed through a subvastus approach under general anesthesia with tourniquet inflation to 300 mmHg. All components were fixated with bone cement. The drainage tube was removed 2 days after operation. Postoperatively, none of the patients did continuous passive motion exercises, but they were encouraged to perform active exercises according to our protocol of rehabilitation. Quadriceps-strengthening exercises were started immediately after surgery and patients began walking with use of a walker on the first postoperative day. On the second postoperative day, the patients started active ROM exercises under the guidance of the medical team.

Clinical and radiographic data were evaluated at postoperative 6 weeks, 3 months, 6 months, and 1 year and then yearly thereafter. Each knee was evaluated according to the Knee Society scoring system (KSS)16) and the Western Ontario and McMaster Universities Osteoarthritis Index (WOMAC)17) by one of the authors (Kim). At the time of each follow-up, active ROM of the knee was measured using a standard 60-cm goniometer with the patient placed in supine position by one of the authors (Kim) who was blinded to the implant used. All complications were recorded. Weight-bearing anteroposterior, lateral, and merchant radiographs were taken at each follow-up visit. All radiographs were reviewed to assess implant position and radiolucent lines. Radiolucent lines were examined at different zones around the femoral and tibial components using the Knee Society TKA radiographic evaluation and scoring system18). All radiolucent lines were recorded and the radiographic sign of loosening was defined as the presence of a radiolucent line more than 2 mm in width at the cement-bone or cement-prosthesis interface19). Implant position in relation to the limb anatomic axis and component alignment was evaluated and the presence and location of the radiolucent lines were assessed using the KSS18). Preoperative and postoperative posterior femoral condylar offsets were also compared.

Preoperative and postoperative clinical outcomes (KSS and WOMAC scores) and radiographic alignment were compared using the paired t-test. A p-value of <0.05 was considered significant, and the analysis was performed using the IBM SPSS ver. 21.0 (IBM Co., Armonk, NY, USA).

Results

The ROM at the final follow-up increased significantly compared to the preoperative ROM (Table 2). The mean KSS improved from 121.4 (range, 42 to 185) preoperatively to 174.0 (range, 130 to 200) at the last follow-up (p<0.001). The mean WOMAC score also improved from 56.1 (range, 23 to 88) preoperatively to 16.4 (range, 0 to 85) postoperatively (p<0.001) (Table 3). The mean KSS and WOMAC scores were 165.8 (range, 105 to 200) and 28.2 (range, 2 to 65), respectively, at the 6-month follow-up, 178.9 (range, 130 to 200 ) and 22.8 (range, 3 to 57), respectively, at the 1-year follow-up, 173.6 (range, 130 to 200) and 17.9 (range, 0 to 85), respectively, at the 2-year follow-up, and 174.2 (range, 130 to 200) and 14.8 (range, 0 to 85), respectively, at the 3-year follow-up.

The mean femorotibial angle was a varus of 3.7°±4.8° (range, varus of 16.8° to valgus of 12.3°) preoperatively and a valgus of 5.7°±1.8° (range, valgus of 0° to valgus of 8.4°) postoperatively (p<0.001). The coronal angle of the femoral component (α) was 96.6°±1.6° (range, 92.1° to 99.9°), the coronal angle of the tibial component (β) was 89.7°±1.7° (range, 84.2° to 95.9°), the flexion angle of the femoral component (γ) was 2.8°±1.3° (range, 0.1° to 5.5°), and the tibial slope angle (δ) was 85.8°±2.1° (range, 82.1° to 90.0°) postoperatively. The mean posterior femoral condylar offset was 29.6±2.5 mm (range, 19.8 to 36.8 mm) preoperatively and 29.9±2.8 mm (range, 20.5 to 36.9 mm) postoperatively (p=0.073). Radiolucent lines were found in 15 knees (14 patients, 7.9%): two in the medial tibial condyle (zones 1 and 2), 2 in the anterior femoral condyle (zones 1 and 2), and 11 in the posterior femoral condyle (zones 3 and 4) (Fig. 2). No case of osteolysis was detected during the 2- to 4-year follow-up period. None of the knees developed aseptic loosening of the femoral, tibial, or patellar component. No knee had a tibiofemoral or patellofemoral joint dislocation.

Two major complications required surgery in two knees. A 70-year-old female with a history of lymphedema developed a deep infection in the extremity of the operated side 1 year postoperatively. The patient had received a radical hysterectomy with total lymph node dissection for uterine cervical cancer. Methicillin-sensitive Staphylococcus aureus was cultured from the joint fluid, and a two-stage revision operation was performed for the infected joint. A patellar fracture occurred in one patient and was treated with open reduction and internal fixation. Three minor complications occurred in three knees. One patient received brisement due to <90° ROM limitation at 4 weeks after the operation and the final flexion angle was 125°. Two patients demonstrated wound dehiscence 2 weeks postoperatively. No infection was detected, and the wound was approximated with simple sutures. Symptomatic deep vein thrombosis, pulmonary embolism, and hyperextension (recurvatum) were not observed during the follow-up.

Discussion

The purpose of this study was to evaluate the ROM and early stability following TKA performed using the new high-flexion prosthesis during a short-term follow-up. The most important finding of the present study was that the newly introduced high-flexion LOSPA prosthesis resulted in no early aseptic loosening of the component and provided outcomes comparable to those used in high-flexion TKA in Korean patients (Table 4).

Postoperative ROM is an important factor for patient satisfaction after TKA, particularly among Asians20). Thus, manufacturers have modified and upgraded the design of knee prostheses to allow for high flexion after TKA. Many studies have investigated the efficacy and safety of TKA using a high-flexion design prosthesis performed by Korean surgeons71113142122232425). However, there has been some puzzling inconsistency among studies on the efficacy of high-flexion TKA designs471112). There is still a paucity of data on the ROM of the knee following TKA using this new high-flexion implant; therefore, we investigated postoperative ROM in patients who received the newly designed high-flexion prosthesis. The LOSPA knee system demonstrated improved clinical and radiographic outcomes at 2- to 4-year follow-up. The outcomes of our study were comparable to those reported by other Korean studies on high-flexion TKA. Seo et al.26) compared the LOSPA PS total knee prosthesis with another high-flexion PS prosthesis. In their study, the mean postoperative flexion angle of the LOSPA group was 126.9°. The mean postoperative ROM of the patients was similar to that of the current study. Theoretically, a knee flexion angle of 140° is possible with the high-flexion LOSPA system, but it appears not to be attainable in vivo.

In the present study, the postoperative ROM of the knee was less than our expectation (up to 140°) and the superiority of the prosthesis compared with conventional TKA in terms of ROM was not observed71122). Several factors may contribute to the ROM after TKA including patient factors (preoperative ROM, obesity, postoperative quality of rehabilitation, patient motivation, and preoperative mental illness), intraoperative technical issues (resection of back osteophytes, ligament balancing, component positioning, altered joint line, and posterior tibial slope), and perioperative complications27). It has been generally considered that the preoperative ROM is the most predictable factor for the postoperative ROM28). Our study showed that the new high-flexion LOSPA prosthesis would not improve the postoperative ROM in patients with a preoperative ROM of <120°. A controversy still exists regarding the relationship between the posterior condylar offset and the ROM629). Bellemans et al.6) reported that the final knee flexion angle was significantly correlated with the posterior femoral condylar offset in TKA patients. They insisted that restoring posterior condylar offset would be very important to achieve a greater flexion angle before impingement of the posterior tibial insert against the posterior femur. In contrast, Hanratty et al.29) found no correlation between the range of knee flexion and posterior condylar offset after TKA. In our study, postoperative posterior condylar offset was well restored compared with the preoperative value. An additional 2-mm posterior femoral bone resection was not necessary because of the 2 mm extension of the posterior condyle of the femoral component. Consequentially, there was no difference in the postoperative posterior condylar offset. Therefore, we found it difficult to determine the relationship between the posterior condylar offset and knee the flexion angle in the present study.

Concerns have been raised about aseptic loosening of the high-flexion knee systems during the early postoperative phase. Cho et al.13) reported that progressive radiolucent lines around the femoral component were detected in 13.8% of 218 knees that received the NexGen LPS-flex TKA at a mean follow-up of 51 months. In their study, radiolucent lines were observed around the anterior and posterior flanges of the femoral component. Han et al.14) noted aseptic loosening in 38% of cases at a mean follow-up of 32 months after using the same high-flexion total knee prosthesis and performed revision in 21% of cases at a mean of 23 months postoperatively. They reported radiolucent lines were most common in the anterior flange of the femoral component. In high-flexion TKA, the tibial insert only articulates with the femoral component, resulting in the absence of load sharing effect and application of greater shearing force on the femoral component. Conventional TKA requires a distinctly higher loosening force compared to high-flexion TKA. Among the high-flexion designs, those with an open geometry of the internal femoral component are more vulnerable to loosening than those with a closed geometry. Peg fixation is essential to good fixation of a component and provides stability against the loosening force30). The LOSPA total knee system requires a 10-mm posterior femoral condylar bone cut, which is more conservative compared to the 12 mm required by the Nexgen LPS-flex system. In addition, peg fixation of the LOSPA prosthesis provides resistance to the loosening of component, but open femoral component geometry is not beneficial for prevention of loosening30). Our results showed a 7.8% incidence of radiolucent lines around the anterior flange of the femur, the posterior femoral condyle, and the medial tibial condylar area. However, no symptoms or signs of aseptic loosening were detected in patients with radiolucent lines. Han et al.14) reported that the mean flexion angle was 136° in the loosening group and 125° in the well- fixed group. The proportion of patients who were able to perform high-flexion activities such as squatting and kneeling was also higher in the loosening group. The femur contacted with the posterior end of tibia at 133° of flexion angle. In our study, the mean flexion angle of the knee was 127°, which was similar with the angle of the well-fixed group and below 133°14). We did not thoroughly investigate the percentage of patients who were able to perform high flexion activities, but we recommended the patients not to be involved in high-flexion activities as much as possible. It was an attempt to prevent early loosening of component following high-flexion TKA13). The knees with appearance of radiolucent lines require further follow-up to assure the clinical course.

Our study has some limitations. First, this was a case series, not a comparative study. We did not have a control group to compare the knee ROM and radiographic and clinical outcomes. Second, the minimum 2-year follow-up period was relatively short, so we cannot determine the long-term clinical and radiographic outcomes.

Conclusions

The evaluation of patients who underwent TKA with the high-flexion LOSPA prosthesis revealed improved clinical and radiographic results and low complication rates after a minimum 2-year follow-up. However, a long-term follow-up is required to determine whether the new high-flexion concept design demonstrates a comparable survivorship. In conclusion, the new high-flexion total knee prosthesis resulted in no early aseptic loosening of the component and improved ROM comparable to other high-flexion TKA prostheses at 2- to 4-year follow-up.

Figures
Fig. 1. Femoral component, tibial insert, and tibial component of the LOSPA total knee prosthesis. LOSPA prosthesis requires removal of additional bone from the posterior femoral condyle to add 10 mm posterior condyle at a large posterior radius of the femoral component. In addition, the femoral component has a more rounded contour and a deepened patellar groove to help deep flexion. The posterior surface of the insert is released for deep flexion favoring design.
Fig. 2. The lateral radiograph of the left knee of a 68-year-old female shows radiolucent lines around the anterior cortex and posterior condylar area 3 years postoperatively. Arrow: these findings were not progressive and the patient had no symptoms or signs of aseptic loosening.
Tables
Patient Demographics

Values are presented as mean±standard deviation or number (range).

TKA: total knee arthroplasty, R: right, L: left, HKA: hip-knee-ankle.

a)Data are presented as femorotibial angle.


Preoperative (Preop) and Postoperative (Postop) Range of Motion

Values are presented as mean±standard deviation (range). p-value<0.001.


Preoperative (Preop) and Postoperative (Postop) Clinical Scores

Values are presented as mean±standard deviation (range). p-value<0.001.

WOMAC: Western Ontario and McMaster Universities.


Summary of Previous Studies on Flexion Angle, Radiolucent Lines, and Aseptic Loosening after Total Knee Arthroplasty Using High Flexion Design Prostheses

F/U: follow-up, Preop: preoperative, Postop: postoperative, NR: no report.


References
  1. Heiberg, K, Bruun-Olsen, V, Mengshoel, AM. Pain and recovery of physical functioning nine months after total knee arthroplasty. J Rehabil Med, 2010;42;614-619.
    Pubmed
  2. Kurosaka, M, Yoshiya, S, Mizuno, K, Yamamoto, T. Maximizing flexion after total knee arthroplasty: the need and the pitfalls. J Arthroplasty, 2002;17;59-62.
    Pubmed
  3. Mulholland, SJ, Wyss, UP. Activities of daily living in non-Western cultures: range of motion requirements for hip and knee joint implants. Int J Rehabil Res, 2001;24;191-198.
    Pubmed
  4. Gupta, SK, Ranawat, AS, Shah, V, Zikria, BA, Zikria, JF, Ranawat, CS. The P.F.C. sigma RP-F TKA designed for improved performance: a matched-pair study. Orthopedics, 2006;29;S49-S52.
    Pubmed
  5. Jones, RE. High-flexion rotating-platform knees: rationale, design, and patient selection. Orthopedics, 2006;29;S76-S79.
    Pubmed
  6. Bellemans, J, Banks, S, Victor, J, Vandenneucker, H, Moemans, A. Fluoroscopic analysis of the kinematics of deep flexion in total knee arthroplasty: influence of posterior condylar offset. J Bone Joint Surg Br, 2002;84;50-53.
    Pubmed
  7. Kim, YH, Sohn, KS, Kim, JS. Range of motion of standard and high-flexion posterior stabilized total knee prostheses: a prospective, randomized study. J Bone Joint Surg Am, 2005;87;1470-1475.
    Pubmed
  8. Argenson, JN, Scuderi, GR, Komistek, RD, Scott, WN, Kelly, MA, Aubaniac, JM. In vivo kinematic evaluation and design considerations related to high flexion in total knee arthroplasty. J Biomech, 2005;38;277-284.
    Pubmed
  9. Jain, S, Pathak, AC, Kanniyan, K, Kulkarni, S, Tawar, S, Mane, P. High-flexion posterior-stabilized total knee prosthesis: is it worth the hype?. Knee Surg Relat Res, 2013;25;100-105.
    Pubmed
  10. Wang, H, Simpson, KJ, Ferrara, MS, Chamnongkich, S, Kinsey, T, Mahoney, OM. Biomechanical differences exhibited during sit-to-stand between total knee arthroplasty designs of varying radii. J Arthroplasty, 2006;21;1193-1199.
    Pubmed
  11. Bin, SI, Nam, TS. Early results of high-flex total knee arthroplasty: comparison study at 1 year after surgery. Knee Surg Sports Traumatol Arthrosc, 2007;15;350-355.
    Pubmed
  12. McCalden, RW, MacDonald, SJ, Bourne, RB, Marr, JT. A randomized controlled trial comparing "high-flex" vs "standard" posterior cruciate substituting polyethylene tibial inserts in total knee arthroplasty. J Arthroplasty, 2009;24;33-38.
    Pubmed
  13. Cho, SD, Youm, YS, Park, KB. Three- to six-year follow-up results after high-flexion total knee arthroplasty: can we allow passive deep knee bending?. Knee Surg Sports Traumatol Arthrosc, 2011;19;899-903.
    Pubmed
  14. Han, HS, Kang, SB, Yoon, KS. High incidence of loosening of the femoral component in legacy posterior stabilised-flex total knee replacement. J Bone Joint Surg Br, 2007;89;1457-1461.
    Pubmed
  15. Kellgren, JH, Lawrence, JS. Radiological assessment of osteoarthrosis. Ann Rheum Dis, 1957;16;494-502.
    Pubmed
  16. Insall, JN, Dorr, LD, Scott, RD, Scott, WN. Rationale of the Knee Society clinical rating system. Clin Orthop Relat Res, 1989;;13-14.
    Pubmed
  17. Bellamy, N, Buchanan, WW, Goldsmith, CH, Campbell, J, Stitt, LW. Validation study of WOMAC: a health status instrument for measuring clinically important patient relevant outcomes to antirheumatic drug therapy in patients with osteoarthritis of the hip or knee. J Rheumatol, 1988;15;1833-1840.
    Pubmed
  18. Ewald, FC. The Knee Society total knee arthroplasty roentgenographic evaluation and scoring system. Clin Orthop Relat Res, 1989;;9-12.
    Pubmed
  19. Gonzalez, MH, Mekhail, AO. The failed total knee arthroplasty: evaluation and etiology. J Am Acad Orthop Surg, 2004;12;436-446.
    Pubmed
  20. Park, KK, Shin, KS, Chang, CB, Kim, SJ, Kim, TK. Functional disabilities and issues of concern in female Asian patients before TKA. Clin Orthop Relat Res, 2007;461;143-152.
    Pubmed
  21. Han, HS, Kang, SB. Brief followup report: does high-flexion total knee arthroplasty allow deep flexion safely in Asian patients?. Clin Orthop Relat Res, 2013;471;1492-1497.
    Pubmed
  22. Choi, WC, Lee, S, Seong, SC, Jung, JH, Lee, MC. Comparison between standard and high-flexion posterior-stabilized rotating-platform mobile-bearing total knee arthroplasties: a randomized controlled study. J Bone Joint Surg Am, 2010;92;2634-2642.
    Pubmed
  23. Seon, JK, Park, JK, Shin, YJ, Seo, HY, Lee, KB, Song, EK. Comparisons of kinematics and range of motion in high-flexion total knee arthroplasty: cruciate retaining vs. substituting designs. Knee Surg Sports Traumatol Arthrosc, 2011;19;2016-2022.
    Pubmed
  24. Lee, BS, Chung, JW, Kim, JM, Kim, KA, Bin, SI. High-flexion prosthesis improves function of TKA in Asian patients without decreasing early survivorship. Clin Orthop Relat Res, 2013;471;1504-1511.
    Pubmed
  25. Kim, TH, Lee, DH, Bin, SI. The NexGen LPS-flex to the knee prosthesis at a minimum of three years. J Bone Joint Surg Br, 2008;90;1304-1310.
    Pubmed
  26. Seo, JG, Moon, YW, Chang, MJ, Jo, BC, Park, YB, Lim, DS, Lee, BH. Design modifications of high-flexion TKA do not improve short term clinical and radiographic outcomes. BMC Musculoskelet Disord, 2014;15;433.
    Pubmed
  27. Dennis, DA, Komistek, RD, Scuderi, GR, Zingde, S. Factors affecting flexion after total knee arthroplasty. Clin Orthop Relat Res, 2007;464;53-60.
    Pubmed
  28. Ritter, MA, Harty, LD, Davis, KE, Meding, JB, Berend, ME. Predicting range of motion after total knee arthroplasty. Clustering, log-linear regression, and regression tree analysis. J Bone Joint Surg Am, 2003;85-A;1278-1285.
    Pubmed
  29. Hanratty, BM, Thompson, NW, Wilson, RK, Beverland, DE. The influence of posterior condylar offset on knee flexion after total knee replacement using a cruciate-sacrificing mobile-bearing implant. J Bone Joint Surg Br, 2007;89;915-918.
    Pubmed
  30. Bollars, P, Luyckx, JP, Innocenti, B, Labey, L, Victor, J, Bellemans, J. Femoral component loosening in high-flexion total knee replacement: an in vitro comparison of high-flexion versus conventional designs. J Bone Joint Surg Br, 2011;93;1355-1361.
    Pubmed


e-submission

This Article


Cited By Articles

Social Network Service

Archives

Indexed/Covered by